Measured and Simulated Performance of a Four-Segmented Clover Detector
基本信息来源于合作网站,原文需代理用户跳转至来源网站获取
摘要:
Gamma rays with energies between 80keV and 1500keV produced by 133Ba and 152Eu standard sources are measured with the clover detector.Relative efficiencies and add-back factors are determined in both crystals and full clover modes.The add-back characteristics have been calculated using Monte-Carlo simulation code Geant4.The simulation and measurements agree very well.