基本信息来源于合作网站,原文需代理用户跳转至来源网站获取       
摘要:
A digital biomedical electrical impedance tomography (EIT) system is developed with the aid of FPGA. The key elements of EIT system are described specifically in the paper. The functions are realized to generate excitation source, switch electrode channels, deal collected signals, demodulate measured voltages etc. The system is tested by a circular tank with 16 stainless electrodes attached around the boundary. The adjacent incentive adjacent measurement mode is adapted to collect boundary voltages of the interesting field. By testing, the system works with 36 dB signal-to-noise ratio (SNR) when 1 mA 100 KHz current is applied into a homogenous tank.