摘要:
Purpose: To assess the possibility of using a public calibration function for radiochromic film dosimetry in dose QA of highly conformal treatment plans. Methods: EBT3 film calibration strips (3.5 × 20 cm2 from lots A101212 and A011713) were exposed on a Varian Trilogy at a facility to a 10 × 10 cm2 open field at doses of 80, 160, 320 cGy using 6MV photons. Together with a strip of unexposed film from the same lot, the exposed films were digitized in a single scan using different Epson 10,000 XL scanners at two different facilities. The dose-response data for each color-channel from each facility were generated using the same calibration function X(D) = a + b/(D - c), where X(D) is the response at dose D and a, b and c are the coefficients. Different batches of EBT3 film were exposed to a VMAT beam. These films, plus two reference strips exposed to doses of zero and 160 cGy, were digitized on the scanners at the two facilities. Using the multi-channel dosimetry method and One-scan protocol (Med Phys, 39: 6339-6349, 2012) the recorded doses on the VMAT films were calculated and the results were compared with the VMAT plan using a Gamma index of 3%/3 mm. Results: The passing rates obtained for dose maps calculated for all combinations of VMAT images and calibration functions were nearly unchanged, using the One-scan protocol. Also, in all cases a passing rate of >99% was obtained for Gamma index of 3%/3 mm. On the other hand, if the One-scan protocol was not employed, the dose maps for VMAT images and calibration functions from different scanners showed poor correlation with the treatment plan. This is probably due to the scan-to-scan variability. Conclusions: The authors have found that it is feasible to use a public calibration function for a given radiochromic film lot using the same methodology, One-scan protocol, for patient-specific QA.
推荐文章
基于Internet的数字示波器e-Calibration技术研究
远程校准
数字存储示波器
互联网
基于加权Ripley's K-function的多尺度景观格局分析——以江苏盐城滨海湿地为例
盐城滨海湿地
景观空间格局
多尺度空间聚集分析
加权Ripley's K-function
空间特征尺度
空间分布强度
利用E-film软件实现放射影像的DVD存档
E-film软件
影像资料存档
DICOM3.0
影响储存与传输系统
内容分析
关键词云
关键词热度
相关文献总数  
(/次)
(/年)
文献信息
篇名 Is It Possible to Publish a Calibration Function for Radiochromic Film?
来源期刊 医学物理学、临床工程、放射肿瘤学(英文) 学科 医学
关键词 DOSIMETRY Quality ASSURANCE IMRT Radiochromic Film MULTICHANNEL
年,卷(期) 2014,(1) 所属期刊栏目
研究方向 页码范围 25-30
页数 6页 分类号 R73
字数 语种
DOI
五维指标
传播情况
(/次)
(/年)
引文网络
引文网络
二级参考文献  (0)
共引文献  (0)
参考文献  (0)
节点文献
引证文献  (0)
同被引文献  (0)
二级引证文献  (0)
2014(0)
  • 参考文献(0)
  • 二级参考文献(0)
  • 引证文献(0)
  • 二级引证文献(0)
研究主题发展历程
节点文献
DOSIMETRY
Quality
ASSURANCE
IMRT
Radiochromic
Film
MULTICHANNEL
研究起点
研究来源
研究分支
研究去脉
引文网络交叉学科
相关学者/机构
期刊影响力
医学物理学、临床工程、放射肿瘤学(英文)
季刊
2168-5436
武汉市江夏区汤逊湖北路38号光谷总部空间
出版文献量(篇)
236
总下载数(次)
0
总被引数(次)
0
论文1v1指导