This work investigated the absorbed dose to water rate under reference conditions in a Cyberknife VSI system using radiochromic films EBT3 and MD-V3 and three ionization chambers: an Exradin A12 and two FC65P Welhöfer Scanditronix with different serial numbers. The correction factor,, was studied using a Varian iX linac and the Cyberknife system. The measurements in the Varian iX were performed in a 10 × 10 cm2 field, 10 cm depth in liquid water at 90 cm and 70 cm SSD and in a 5.4 × 5.4 cm2 field, 10 cm depth at 70 cm SSD to simulate the Cyberknife conditions. In the Cyberknife system, measurements were performed using ionization chambers and both film types at 70 cm SSD and 10 cm depth in its 6 cm diameter reference field. The results indicate that ?is independent of the dosimeters and the evaluation methods. Maximum differences of 0.22% - 0.55% (combined uncertainties of 1.22% - 1.98%, k = 1) are obtained on ?using Varian iX, whereas discrepancies of 2.08% - 2.09% (combined uncertainties of 1.87% - 2.13%, k = 1) are observed using the Cyberknife system. Given the agreement between detectors and the combined standard uncertainties, the data from Varian iX could be considered the most accurate and consequently a weighted average factor of 0.902 ± 0.006 could be used for the Cyberknife VSI system reference field. Within measurement uncertainties, the absorbed dose rate measured in the Cyberknife VSI system reference field was found to be independent of the dosimeters used. These results suggest that the absorbed dose measured at a point within a given field size should be the same, regardless the dosimeter used, if their dosimetric characteristics are well known. This highlighted the importance of performing dosimetry by controlling all parameters that could affect the dosimeter response. One can conclude that radiochromic film dosimetry can be considered as an appropriate alternative for measuring absorbed dose to water rate.
推荐文章
Groundwater quality assessment using multivariate analysis, geostatistical modeling, and water quali
Groundwater
Multivariate analysis
Geostatistical modeling
Geochemical modeling
Mineralization
Ordinary Kriging
Determination of brominated diphenyl ethers in atmospheric particulate matter using selective pressu
Brominated diphenyl ethers
Atmospheric particulate matters
Selective pressurised liquid extraction
Gas chromatography-mass spectrometry
内容分析
关键词云
关键词热度
相关文献总数  
(/次)
(/年)
文献信息
篇名 Absorbed Dose to Water Rate in a Cyberknife VSI System Reference Field Using Ionization Chambers and Gafchromic Films
来源期刊 医学物理学、临床工程、放射肿瘤学(英文) 学科 医学
关键词 Small FIELD DOSIMETRY Reference FIELD CYBERKNIFE Gafchromic Films Ionization CHAMBERS Dose RATE EBT3 MD-V3 6 MV X-Rays
年,卷(期) 2017,(1) 所属期刊栏目
研究方向 页码范围 80-92
页数 13页 分类号 R73
字数 语种
DOI
五维指标
传播情况
(/次)
(/年)
引文网络
引文网络
二级参考文献  (0)
共引文献  (0)
参考文献  (0)
节点文献
引证文献  (0)
同被引文献  (0)
二级引证文献  (0)
2017(0)
  • 参考文献(0)
  • 二级参考文献(0)
  • 引证文献(0)
  • 二级引证文献(0)
研究主题发展历程
节点文献
Small
FIELD
DOSIMETRY
Reference
FIELD
CYBERKNIFE
Gafchromic
Films
Ionization
CHAMBERS
Dose
RATE
EBT3
MD-V3
6
MV
X-Rays
研究起点
研究来源
研究分支
研究去脉
引文网络交叉学科
相关学者/机构
期刊影响力
医学物理学、临床工程、放射肿瘤学(英文)
季刊
2168-5436
武汉市江夏区汤逊湖北路38号光谷总部空间
出版文献量(篇)
236
总下载数(次)
0
总被引数(次)
0
论文1v1指导