Purpose: to compare the rebound tonometer ICare? (RT) with the Goldmann applanation tonometer (GAT) in cataract surgery and to assess the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements. Design: retrospective, comparative study. Methods: a total of 472 patients underwent IOP measurement using RT (time 0 = RTa). GAT IOP measurement was performed 5 minutes later, followed by a second RT IOP measurement after other 5 minutes (RTa + 10 min = RTb). CCT was obtained by ultrasound pachymetry. In 106 patients IOP was measured by means of RT and GAT before clear corneal cataract surgery (RT1 and GAT1) and at one day postoperatively (RT2 and GAT2). Results: RT IOP values > 5 mmHg were overestimated, while RT IOP values Conclusion: our results showed a good agreement between measurements obtained with RT and GAT in pre and postoperative cataract surgery, although a significant statistically difference was found between RT and GAT measurements performed postoperatively. Moreover, CCT is a parameter to be considered for the IOP measurement, especially for values obtained with RT.
推荐文章
4种眼压计对近视患者准分子激光原位角膜磨镶术后测压的应用价值比较——附53例报告
近视
准分子激光原位角膜磨镶术
中央角膜厚度
压平眼压计
动态轮廓眼压计
4种眼压计对近视患者准分子激光原位角膜磨镶术后测压的应用价值比较——附53例报告
近视
准分子激光原位角膜磨镶术
中央角膜厚度
压平眼压计
动态轮廓眼压计
内容分析
关键词云
关键词热度
相关文献总数  
(/次)
(/年)
文献信息
篇名 Rebound Tonometry in Cataract Surgery: Comparison with Goldmann Applanation Tonometry
来源期刊 眼科学期刊(英文) 学科 医学
关键词 Applanation TONOMETER CATARACT SURGERY INTRAOCULAR Pressure PACHYMETRY REBOUND TONOMETER
年,卷(期) 2018,(1) 所属期刊栏目
研究方向 页码范围 1-11
页数 11页 分类号 R77
字数 语种
DOI
五维指标
传播情况
(/次)
(/年)
引文网络
引文网络
二级参考文献  (0)
共引文献  (0)
参考文献  (0)
节点文献
引证文献  (0)
同被引文献  (0)
二级引证文献  (0)
2018(0)
  • 参考文献(0)
  • 二级参考文献(0)
  • 引证文献(0)
  • 二级引证文献(0)
研究主题发展历程
节点文献
Applanation
TONOMETER
CATARACT
SURGERY
INTRAOCULAR
Pressure
PACHYMETRY
REBOUND
TONOMETER
研究起点
研究来源
研究分支
研究去脉
引文网络交叉学科
相关学者/机构
期刊影响力
眼科学期刊(英文)
季刊
2165-7408
武汉市江夏区汤逊湖北路38号光谷总部空间
出版文献量(篇)
241
总下载数(次)
0
总被引数(次)
0
论文1v1指导