A parametric study was performed to design a device capable of treating small targeted regions within the prostate using high intensity focused ultrasound, while sparing the surrounding organs and minimizing the number of elements. The optimal focal length (L), operating frequency (f), element size (a) and central opening radius for lodging an imaging probe (r) of a device that would safely treat tissue within the prostate were obtained. Images from the Visible Human Project were used to determine simulated organ sizes and treatment locations. Elliptical tumors were placed throughout the simulated prostate and their lateral and axial limits were selected as test locations. Using graphics processors, the acoustic field and Bio-Heat Transfer Equation were solved to calculate the heating produced during a simulated treatment. L, f, a and r were varied from 45 to 75 mm, 2.25 to 3 MHz, 1.5 to 8 times the wavelength and 9 to 12.5 mm, respectively. The resulting optimal device was a 761-element concentric-ring transducer with L = 68 mm, f = 2.75 MHz, a = 2.05λ and r = 9 mm. Simulated thermal lesions showed that it was possible to treat target tumors consistent with reported locations and sizes for prostate cancer.
推荐文章
Thermodynamic properties of San Carlos olivine at high temperature and high pressure
San Carlos olivine
Thermodynamic property
Thermal expansion
Heat capacity
Temperature gradient
An experimental study of interaction between pure water and alkaline feldspar at high temperatures a
Alkaline feldspar
Autoclave
High-temperature and high-pressure experiments
Effects of a proline solution cover on the geochemical and mineralogical characteristics of high-sul
Proline
Coal gangue
Pollution control
Heavy metal fraction
Mineralogical characteristics
Iron isotope fractionation during fenitization: a case study of carbonatite dykes from Bayan Obo, In
Iron isotopes
Fenitization
Fluid exsolution
Carbonatite
Bayan Obo
内容分析
关键词云
关键词热度
相关文献总数  
(/次)
(/年)
文献信息
篇名 High Intensity Focused Ultrasound Optimal Device Design for Targeted Prostate Cancer Treatment: A Numerical Study
来源期刊 现代物理(英文) 学科 医学
关键词 FOCUSED Ultrasound PROSTATE Cancer Thermal Ablation NON-INVASIVE TREATMENT BHTE HIFU
年,卷(期) 2013,(2) 所属期刊栏目
研究方向 页码范围 240-245
页数 6页 分类号 R73
字数 语种
DOI
五维指标
传播情况
(/次)
(/年)
引文网络
引文网络
二级参考文献  (0)
共引文献  (0)
参考文献  (0)
节点文献
引证文献  (0)
同被引文献  (0)
二级引证文献  (0)
2013(0)
  • 参考文献(0)
  • 二级参考文献(0)
  • 引证文献(0)
  • 二级引证文献(0)
研究主题发展历程
节点文献
FOCUSED
Ultrasound
PROSTATE
Cancer
Thermal
Ablation
NON-INVASIVE
TREATMENT
BHTE
HIFU
研究起点
研究来源
研究分支
研究去脉
引文网络交叉学科
相关学者/机构
期刊影响力
现代物理(英文)
月刊
2153-1196
武汉市江夏区汤逊湖北路38号光谷总部空间
出版文献量(篇)
1826
总下载数(次)
0
总被引数(次)
0
论文1v1指导