A New Pumping-Probing Scheme for the Optically Pumped Cesium Beam Frequency Standard
基本信息来源于合作网站,原文需代理用户跳转至来源网站获取
摘要:
A new pumping-probing scheme for the optically pumped cesium beam frequency standard has been experimentally tested in our laboratory. The stability of the optically pumped cesium beam frequency standard was measured by comparing its 10 MHz output with an HP5071A commercial cesium atomic clock. The result shows that the frequency stability for the 1 s and 30000s sample times are 1.2 × 10-11 and 3.7 × 10-13, respectively. It was proved that the new pumping scheme works well.