Determination of Rifampicin by Peroxomonosulfate-Cobalt(Ⅱ)Chemiluminescence System
基本信息来源于合作网站,原文需代理用户跳转至来源网站获取
摘要:
Rifampicin can enhance the chemiluminescence(CL)of peroxomonosulfate-cobalt(Ⅱ)system,and the CL intensity is strongly dependent on the rifampicin concentrations.Based on this phenomenon,a rapid and sensitive flow injection CL method was developed for the determination of rifampicin.The relative CL intensity was linear with the rifampicin concentration over the range of 5×10-8 to 1×10-6 g·mL-1(r=0.9991),the detection limit was 7×1019 g·mL-1(S/N=3),and the relative standard deviation was 2.7%for 6×10-7g·mL-1 rifampicin(n=11).Furthermore,this method Was successfully applied to the determination of rifampicin in real eye drop and capsules sample.